Investigating Aluminum Alloys Using Micro-X-Ray Fluorescence Spectroscopy (Micro-XRF) in SEM and EDS

汽车和飞机工业经常使用铝合金,将合金的机械性能和铝的轻重方面结合在一起。元素的组成对于合金特性至关重要,因为即使是痕量的添加剂也会影响材料的热和机械性能。

Micro-X-ray fluorescence spectroscopy (micro-XRF) in the scanning electron microscope (SEM) used in the study of aluminum alloys is promising because of the higher sensitivity of detection of elements (Z > 20). This article illustrates the improvement of the detection of trace elements in a certified alloy of aluminum (Alcoa Deltalloy®4032,图1.)在SEM上使用micro-XRF

分析的标本Alcoa Deltalloy的照片® 4032.

Figure 1.分析的标本Alcoa Deltalloy的照片®4032.

方法

A focused source of X-ray photon, XTrace, facilitates the performance micro-XRF in a SEM. This dual working system enables SEM to detect trace elements at the parts per million (ppm) level. The X-ray source on the SEM enhances the detectability of elements with an atomic number higher than 20, compared with SEM-EDS. A decreased spectral background is the result of photon excitation as there is no generation of bremsstrahlung in the sample, compared to the excitation of an electron. The data shows a better peak ratio to background ratio. In addition to the possibility to effectively excite the high energy lines over 15 keV, the problems that occur from overlap between neighboring peaks are reduced.

The limit of detection using micro-XRF can be as low as 10 ppm (µg/g). The sample matrix and the element influence the limit of detection (Figure 2). To study comparatively EDS can be used only to detect trace elements that are as low as 1000 pm (0.1 mass%). Generally the detective sensitivity for elements that are heavier than calcium is higher when micro-XRF is used instead of an electron beam (Figure 2). The excitation of X-ray gives information regarding the composition of the sample surface, much like SEM, but it also provides information from a depth of several microns.

Estimated detection limits for photon excitation (red) and electron excitation (blue). They depend on the element and the matrix composition. In general, Micro-XRF improves the detectability of elements above atomic number 20 (calcium) - compared to SEM EDS. Only Micro-XRF detects the elements shown in black, their concentration is below or close to the limit of detection for electron excitation.

Figure 2.Estimated detection limits for photon excitation (red) and electron excitation (blue). They depend on the element and the matrix composition. In general, Micro-XRF improves the detectability of elements above atomic number 20 (calcium) - compared to SEM EDS. Only Micro-XRF detects the elements shown in black, their concentration is below or close to the limit of detection for electron excitation.

Conditions for Measurement

A SEM was integrated with aBruker Quontax系统with a 30 mm² active area silicon drift detector (SDD) XFlash®6I30 with energy resolution of 123 eV for Mn Ka and XTrace source with an Rh-anode X-ray tube. Electron excitation was performed using SEM, and photon excitation was completed using XTrace. X-rays that were generated from both of these sources, electron excitation and photon excitation, were detected using XFlash®SDD.

该分析需要10点频谱,使用连续和独立的光子和电子激发在样品的同一区域上的每种技术。XRF的定量使用了基本的无标准算法参数。使用峰到背景的定量方法(P/B-ZAF)获得EDS定量的结果。表1给出了使用两种激发方法检测到的测量条件和元素的摘要。

表格1。测量条件和检测到的铝合金中的元素。仅用微XRF检测到红色的元素。

Excitation Voltage, current 输入计数率(ICR) Acquisition time Detected elements
电子(SEM) 20 kV 10.7 kcps 200 s mg,al,si,Cr, Fe, Ni, Cu, Zn
Photons (XTrace) 50 kV, 600 µA 9.5 kcps 200 s mg,al,si,Ti,V,Cr,,Mn, Fe, Ni, Cu, Zn,Ga,Sr

结果

The elements such as Al, Mg, Cr, Si, Ni, Fe, Zn and Cu were present in the certified alloy of aluminum using EDS as well as micro-XRF (Table 1, Figure 3). Other trace elements were present in the sample. Ti, Mn, V, Sr and Ga were the elements that were exclusively isolated using micro-XRF. The ratio of signal to noise is much better with lower detection levels of micro-XRF (Figure 2), facilitating the detection of Fe, Cr, Cu, Ni and Zn.

Comparison of EDS (blue) and Micro-XRF (red) spectrum. The element peaks of Ti, V, Mn, Ga and Sr are only present in the Micro-XRF spectrum. The Rh-lines derived from the scattered XRF-tube spectrum. Unlabeled peaks are diffraction peaks.

Figure 3.Comparison of EDS (blue) and Micro-XRF (red) spectrum. The element peaks of Ti, V, Mn, Ga and Sr are only present in the Micro-XRF spectrum. The Rh-lines derived from the scattered XRF-tube spectrum. Unlabeled peaks are diffraction peaks.

表2中显示了XRD和EDS的无标准定量结果。仅使用EDS对Al,Mg,Fe,Si,Zn,Cu和Ni的量化是可能的。Zn的量化只能在10次运行中的4个中进行,因为它在EDS的其他六个分析运行中的检测极限之下。

表2。Comparison of the EDS and Micro-XRF quantification results with the certified values, element concentration in mass% normalized (b.d. = below limit of detection, n.d. = not detectable)

毫克 Al Si Ti V Cr Mn Fe Cu Zn Ga Sr
EDS mean concentration value 1.20 85.09 11.43 N.D. N.D. B.D. N.D. 0.36 0.91 0.91 0.10 N.D. N.D.
微XRF平均浓度值 0.85 83.87 12.83 0.03 0.01 0.06 0.03 0.36 0.93 0.93 0.10 0.01 0.03
Certificied values 1.1 84.52 12.00 0.011 0.0099 0.051 0.033 0.31 0.89 0.89 0.098 0.02 0.026

The quantification of the micro-XRF spectrum was achieved with a standardless fundamental parameter method. The results of this quantification illustrate the element sensitivity of XTrace. The elements V, Ti, Mn, Cr, Sr and Ga were detectable and quantified using micro-XRF, even though the element concentration was below the detection limit of EDS of 0.1 mass%.

Though chromium was detected with EDS, it could not be quantified using this method as its concentration was below the detection limit of EDS. In contrast, chromium can be detected and quantified using micro-XRF, due to its better detection limit. The comparison of the results of quantification obtained from micro-XRF and EDS to the aluminum alloy’s certified values is illustrated in Table 2. It is observed that the results of quantification for most elements using XRF and EDS are in close proximity to that of the certified values.

A diffraction peak overlap caused a slightly larger variation for Ti concentration between the certified value and micro-XRF result.

结论

Due to its high element sensitivity, five additional elements such as V, Ti, Mn, Sr and Ga were detected exclusively in a certified alloy of aluminum using Micro-XRF.Micro-XRFhigh sensitivity for element detection is due to a lower background, lack of bremsstrahlung, and a peak that is higher to the background ratio when compared to electron excitation, so better sample characterization is possible with micro-XRF. The results of standardless quantification using EDS and Micro-XRF are close to that of the certified values. XTrace serves as a complementary analytical tool for a SEM to enhance the detectability of trace elements.

This information has been sourced, reviewed and adapted from materials provided by Bruker Nano Analytics.

For more information on this source, please visitBruker Nano Analytics.

引用

Please use one of the following formats to cite this article in your essay, paper or report:

  • APA

    Bruker Nano Analytics. (2019, August 29). Investigating Aluminum Alloys Using Micro-X-Ray Fluorescence Spectroscopy (Micro-XRF) in SEM and EDS. AZoM. Retrieved on November 23, 2021 from //www.washintong.com/article.aspx?ArticleID=12675.

  • MLA

    Bruker Nano Analytics. "Investigating Aluminum Alloys Using Micro-X-Ray Fluorescence Spectroscopy (Micro-XRF) in SEM and EDS".azom. 23 November 2021. .

  • 芝加哥

    Bruker Nano Analytics. "Investigating Aluminum Alloys Using Micro-X-Ray Fluorescence Spectroscopy (Micro-XRF) in SEM and EDS". AZoM. //www.washintong.com/article.aspx?ArticleID=12675. (accessed November 23, 2021).

  • Harvard

    Bruker Nano Analytics. 2019.Investigating Aluminum Alloys Using Micro-X-Ray Fluorescence Spectroscopy (Micro-XRF) in SEM and EDS. AZoM, viewed 23 November 2021, //www.washintong.com/article.aspx?ArticleID=12675.

Ask A Question

Do you have a question you'd like to ask regarding this article?

Leave your feedback
Submit